Main technical data
Technical parameters of Anaesthetic Workstation
O2 supply range:
Rated working pressure: 0.4Mpa, allowable range: (0.27~0.55)
Mpa, from central supply or gas cylinder.
Adjusting range of flow: O2: (0.1~10) L/min, N2O: (0.1~10) L/min
Breathing circuit: Working mode: closed, semi-closed (1~4) kpa, semi-opened;
The resistances of inspiration and expiration are all not more than 0.6kpa; CO2 absorb cylinder: 2 pieces,
can load with 1600g Na2CO3 at one time.
Evaporator:
Concentration: 0~5vol% Temperature: (15~35) ° Flow: (0.25~15) L/min
With flow, temperature and pressure compensated automatically, and can select Enflurane or Isoflurane according
to users' needs
Alarm for O2 supply failure and N2O automatic cut-off device:
Audible alarm will be given for at least 7 seconds when O2 supply pressure drops below 0.2Mpa; gas cut-off
device makes the flow of N2O stop before that of O2 when O2 supply pressure is lower than alarm pressure
and is still falling.
Technical parameters of Anaesthetic Ventilator
Tidal volume: (50~1500) mL
Frequency: (6~60) times/min
I: E: 1:1~1:4.
Inspiratory flow: (0.1~1.0) L/s
Airway pressure: (-2~10) Kpa
Upper limit pressure (include limit) : (1~6) Kpa
Lower limit pressure: (0.4~4) kpa, the alarm will be given for 4s~15s when it drops below the setting value
of lower limit.
Bellows: Adult :(300~1500) mL; Child:(50~300) mL
Category :
Other Medical Equipment
Related Searches :
Country Of Origin :
China
Listing Type :
Sell
Brand Name :
EXPANDING
Model Number :
EX-941
Minimum Order Quantity :
1
Warranty :
one year fo rmachine,6 months for accessories
Certifications :
ISO9001,ISO13485
Contact :
Reikee Mok
Address :
5A, Tower 3, Hua Cai Tian Chen
Wen Xin 2nd Road, Nanshan Cen
Shenzhen, Guangdong, 518045
China
Telephone :
[86] (755) 26422457
Fax :
[86] (755) 26422562