Sell AR-100A Medical Ventilator

AR-100A Medical Ventilator You May Also Be Interested In: medical ventilator pressure indicator pressure reducer simulators ventilator
Physical Specifications of medical ventilator
Dimensions Width: 50cm Depth: 50cm Height: 120cm
Weight 20Kg

Environmental Specifications of medical ventilator
Operating Requirements Storage/Shipping Requirements
Temperature: 5 to 40oC Temperature: -20 to 70oC
Relative Humidity: less than 80% Relative Humidity: less than 93%
The atmosphere pressure: 70 to 106 kPa The atmosphere pressure: 70 to 106 kPa

Output Voltage Specification Information of medical ventilator
AC Voltage: 220V+/-10%
Frequency: 50HZ+/-2%
Power: 130VA

Pneumatic Specifications of medical ventilator
Oxygen Inlet
Pressurized Oxygen: 0.28~0.6 MPa medical oxygen

IEC Classification of medical ventilator
Ordinary equipment, not protected against the ingress of liquids Not protected/Not suitable for use in the presence of flammable anesthetic gases
Class "q, Type B

Performance Specification of medical ventilator
Power ON/OFF
Tidal Volume 0~1800ml
Breath Rate 5~60 bpm
%O2 45~90%
Inspiratory Time 0.5~3sec
Inspiratory Pause 0~1sec
Trigger Level -1~0kpa
Over Pressure Relief <6kpa

Alarms and Indicators of medical ventilator
High Pressure Alarm Limit flash, red indicator, can be adjustable
Low Pressure Alarm Limit Red indicator , can be adjustable
High Minute Volume Alarm Limit flash, Red indicator, can be adjustable
Low Minute Volume Alarm Limit flash, Red indicator, can be adjustable
High Rate Alarm Limit flash, Red indicator , can be adjustable
Apnea interval flash, Red indicator, can be adjustable
Alarm Silence 120sec. max
Power Red indicator
Insp. phase Green indicator
Exh. phase Green indicator
T. VOL Indicator Green indicator
MIN. VOL Indicator Green indicator
PRESSURE Indicator Green indicator
RATE Indicator Green indicator

What Is Provided of AR-100A Medical Ventilator
Part Name Number Specification
Ventilator 1
Mobile 1
Humidifier 1
Breath Circuit 1
Manipulators 1
Oxygen Pressure Reducer 1
Mask 1
Lung Simulators 1
Hand Book 1
CD For Operation 1

Electrical Specifications And Fuse Specification Information
Input Voltage Requirements Quantity: 2
AC Voltage: 220V+/-10% Current Rating: 0.8 A
Frequency: 50HZ+/-2% Voltage Rating: 250 Volts
Power: 170 VA Type: 5*20 mm RT