Sell Mobile Surgical X-ray Machine

Mobile Surgical X-ray Machine You May Also Be Interested In: mobile x ray x ray machine
Fluoroscopic Capacity
Max rated capacity:Tube current 3mA, Tube voltage 90kV
Automatic Fluoroscopy
Tube voltage: 45kV~90kV, adjust automatically
Tube current: 0.3mA~3mA, set manually
Manual Fluoroscopy
Tube voltage: 45kV~90kV, in succession
Tube current: 0.3mA~3mA, in succession
Radiography Capacity
Max rated capacity:1.73kVA (70kV, 35mA, 0.1s)
Fixed combination of tube voltage and tube current, classified into 5 grades: 45kV50mA, 60kV40mA, 70kV35mA, 80kV25mA, 90kV20mA
Time:0.1s~7s with an interval of 0.1s
Size of plateholder:200mm * 250mm (8"*10") or 250mm *300mm (10"*12")
X-ray Tube:XD6-1.1, 3.5/100
Fixed anode x-ray tube with 2 focus:large focus 2.6mm, small focus 1mm,
Field of view:The unit applies 150mm (6") EGA made by Toshiba, Japan. And 170mm (7") or 230mm(9") is optional.
Image resolution:Horizontal 800 lines and Vertical 400 line
Gray scale: 8
Resolution: >= 14LP/cm
Line Scanned: 625
Image/sec: 25
Direction wheel Horizontal movement, +/-90 degrees rotation
Ascending & descending range >=400mm
C-arm C-arm moving distance: 200mm
Revolution around horizontal axis +/- 180 degrees
Revolution around vertical axis +/- 12.5 degrees
Slip on orbit 90 degrees
Features1. Clinical Video system with high performance, 8 images storage volume, and two 14 inches high-resolution monitors
2. Automatically track fluoroscopy to make the image brightness and clearness optimum
3. Automatically store the last image of fluoroscopy to advantage diagnosis
Brand Name
ISO9001, CE
Available Colors
Condition of Goods
Lead Time from Order to Shipment
30 days
Model Number
Minimum Order Quantity
Power Requirements
110 or 240 volts, 50/60 hz
Terms of Payment
L/C at sight
Terms of Sale
FOB Shanghai